Evaluation of accuracy and reliability of PulseOn optical heart rate monitoring device
No Thumbnail Available
Author
Delgado-Gonzalo, Ricard
Parak, Jakub
Tarniceriu, Adrian
Renevey, Philippe
Bertschi, Mattia
Korhonen, Ilkka
Abstract
PulseOn is a wrist-worn optical heart rate (HR) monitor based on photoplethysmography. It utilizes multiwavelength technology and optimized sensor geometry to monitor blood flow at different depths of skin tissue, and it dynamically adapts to an optimal measurement depth in different conditions. Movement artefacts are reduced by adaptive movement-cancellation algorithms and optimized mechanics, which stabilize the sensor-to-skin contact. In this paper, we evaluated the accuracy and reliability of PulseOn technology against ECG-derived HR in laboratory conditions during a wide range of physical activities and also during outdoor sports. In addition, we compared the performance to another on-the-shelf consumer product Mio LINK®. The results showed PulseOn reliability (% of time with error <10bpm) of 94.5% with accuracy (100% - mean absolute percentage error) 96.6% as compared to ECG (vs 86.6% and 94.4% for Mio LINK®, correspondingly) during laboratory protocol. Similar or better reliability and accuracy was seen during normal outdoor sports activities. The results show that PulseOn provides reliability and accuracy similar to traditional chest strap ECG HR monitors during cardiovascular exercise.
Publication Reference
2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Milan (Italy), pp. 430-433
Year
2015